BeJUNK12 and BeJUNK18
Bergmann-Junk (STANAG 5A/5B) and Bergmann-Junk-Siebert (STANAG 5C) tests are traditional methods for the quantitative determination of chemical stability of nitrocellulose and other nitrate esters, single and double-based smokeless powders, rocket fuels and nitro compounds. BeJUNK12 and BeJUNK18 testing apparatus present a cost-effective and modern solution for R&D and quality assurance laboratories across the branch. Efficient metal heating block, LCD touch screen interface are combined into standalone apparatus. The device can be connected to a PC for an online acquisition of temperature values and remote operation. Quality glassware and rigid tubes holder improve the operation effectivity.
GENERAL SPECIFICATIONS
BeJUNK12 and BeJUNK18 are modular units which follow 2 basic configurations. Converted is easy and requires no extra heating block.
Typical sample weight | 2 grams |
Hole diameter | 20 mm or as required |
Hole depth deep/reduced | 200 / 150 mm |
Typical testing temperature | 132 (115 and 120) °C |
Typical testing temperature | 2 (8 and 16) hours |
Block temperature limit | + 200 °C |
No. of samples | 12 / 18 |
STANAG 5A / 5B certified glassware
Tube diameter | 19 mm |
Tube length | 350 mm |
Absorber type | Cap |
Typical liquid used | water |
Typical liquid volume | 25 ml |
NC type | chalked / non-chalked |
STANAG 5C certified glassware
Tube diameter | 19 mm |
Tube length | 270 mm |
Absorber type | Globe |
Typical liquid used | Hydrogen peroxide 3% |
Typical liquid volume | 50 ml |
NC type | non-chalked only |
ADVANTAGES
A standalone device with LCD touchscreen interface |
General test timer on the LCD screen |
Sound and light reminder of experiment finish |
Accurate temperature controlling |
Independent overheating protection |
User-defined testing procedures (self-programming) |
External calibration |
Remote operation via USB/LAN |
Firmware update via SD card |
Rigid stands for sample preparation and cooling |
Permanently marked glassware |
COMPLIANCE
STANAG 4178 (5A, 5B, 5C), COS 137602, TL: 1376-0589, TL: 1376-0600, UK M28/89
BeJUNK12 ACCESSORIES
-
No subcategories